1.888.610.7664

sales@GlobalTestSupply.com

Customer Reviews
 
 

Fluke 1523-P4 Reference (IR) Thermometer Kit with carrying case

Model: 1523-P4 | Order No: 1523-P4-156

Measures, graphs, and records PRTs thermocouples, and thermistors, making this reference thermometer ideal for field applications, laboratory measurements, and data logging. Also, it provides high accuracy, a wide measurement range, logging, trending, and comes equipped with a magnetic meter hanger and connectors.

Discontinued!

This product has been discontinued and is no longer available.
This is the direct replacement:

Fluke 1524-P4 Dual-Channel Reference Thermometer Kit with 5615 PRT-

Fluke 1524-P4 Dual-Channel Reference Thermometer Kit with 5615 PRT

Model: 1524-P4 | Order No: 1524-P4-156

The ideal bundle for you to measure, graph, and record PRTs, thermocouples, and thermistors with one device. The dual-channel thermometer readouts deliver; high accuracy, wide measurement range, logging, and trending, all in a handheld tool you can take anywhere.



Fluke 1523-P4 Offers

Measure, graph and record three sensor types with one tool.

Features

  • Confidently cover any temperature from –200 to 420°C with included 5615 platinum resistance thermometer with accredited calibration
  • Protect their investment with sturdy carrying case
  • Plug any thermocouple with mini-thermocouple jacks into the universal thermocouple adapter for convenient measurement
  • Solve almost any positioning problem with TPAK universal magnetic hanging clip

Applications

  • Instrument Technicians
  • Process Technicians
  • Calibration Technicians
  • Researchers in Process Industries
  • Pharmaceutical
  • Biotech Plants
  • Medical Device Manufacturing
  • Power Generation/Transmission
  • Aerospace

Fluke 1523-P4 Specifications

Input Channels 1
Resolution PRTs and thermistors: 0.001°
Thermocouples: 0.01°
Logging 25 readings with statistics
Sample Interval (Normal) 1 second
Typical Sample Interval (Fast Mode) 0.3 seconds
Sensor Types PRTs, RTDs, thermistors, and thermocouples
Thermocouple Types C, E, J, K, L, M, N, T, U, B, R, S
Statistics Maximum, minimum, average, standard deviation
Trending Scale: ±18°F (10°C), ±1.8°F (1°C), ±0.18°F (0.1°C), ±0.018°F (0.01°C), 10 minutes of real-time data
Power Requirements 3 x AA alkaline batteries, 12 V dc universal power supply
Computer Interface RS-232, 9940 I/O ToolKit software included
Environmental Conditions for Best Accuracy 55.4 to 91.4°F (13 to 33°C)
Millivolt Range and Accuracy 10 to 75 mV, ± (0.005% + 5 µV)
Internal Reference Junction Compensation ±0.36°F (±0.2°C)
Resistance Range and Accuracy 0 to 400 Ω ± (0.004% + 0.002 Ω)
200 Ω to 50 kΩ ± (0.01% + 0.5 Ω)
50 to 500 kΩ ± (0.03 %)
Temperature Coefficient, Voltage ±(0.001%/°C + 1 mV/°C)
Temperature Coefficient, Resistance 0.0008%/°C + 0.0004 Ω (0 to 400 Ω)
0.002%/°C + 0.1 Ω (0 Ω to 50 kΩ)
0.06%/°C + 0.1 Ω (50 to 500 kΩ)
Excitation Current, Resistance 1 mA (0 to 400 Ω)
10 µA (0 Ω to 50 kΩ)
2 µA (50 to 500 kΩ)
Dimensions 3.75 x 7.9 x 1.86" (96 x 200 x 47 mm)
Weight 1.45 lbs (0.65 kg)
Safety EN61010-1:2001, CAN/CSA C22.2 No. 61010.1-04
Click here for complete specifications on the Fluke 1523-P4

What's included with the Fluke 1523-P4

  • Thermometer
  • NIST Traceable Certificate of Calibration
  • Connector
  • Magnetic Meter Hanger
  • Cable
  • ToolKit Software
  • Power Supply
  • Users Guide
  • CD-ROM (Contains Technical Manual)

Accessories

Click on a category to view a selection of compatible accessories with the Fluke 1523-P4 Reference (IR) Thermometer Kit with carrying case.

Fluke TB25 Tool Bucket Organizer Bag, 5.2 gal TB25
Durable and rugged, this tool bucket organizer bag has a 5.2-gallon capacity that can be relied on to protect and safely carry a variety of test tools and accessories. With its waterproof hard bottom shell, this versatile organizer has a total of 28 pockets that can fit hand tools, DMMs, clamp meters, test probes, and more.