Model: 5609-300-P
Achieve high-precision measurements over a wide temperature range with this stable secondary PRT, which features INFO-CON connections and works with 1523 and 1524 models. It comes with a 300mm long sheath with a 6mm diameter.
Features
| Temperature Range | -328 to 1,238°F (-200 to 670°C) | 
| Nominal Resistance at 0.01°C | 100Ω ±0.5Ω | 
| Temperature Coefficient | 0.0039250Ω/Ω/°C | 
| Short-Term Repeatability | ±0.01 at 0.010°C; ±0.02°C at max temp | 
| Drift | ±0.01 at 0.010°C; ±0.02°C at max temp | 
| Hysteresis | ±0.01°C maximum | 
| Sensor Length | 1.2in ±0.2in (30mm ±5mm) | 
| Sensor Location | 0.1in ±0.1in (3mm ±1mm from tip) | 
| Minimum Insulation Resistance | 500MΩ at 23°C, 10MΩ at 670°C | 
| Transition Junction Dimensions | 2.8 x 0.49in (71 x 12.5mm) | 
| Minimum Immersion Length (<5 mK Error) | 3.9in (100mm) | 
| Maximum Immersion Length | 12in (305mm) | 
| Response Time | 12 seconds typical | 
| Self Heating (In 0°C Bath) | 50mW/°C | 
| Lead-Wire Cable Type | Teflon, 24AWG | 
| Lead-Wire Length | 6ft (1.8m) | 
| Lead-Wire Temperature Range | -58 to 482°F (-50 to 250°C) | 
| Calibration | Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0; please see calibration uncertainty table and its explanation of changeable uncertainties | 
Click on a category to view a selection of compatible accessories with the Fluke 5609-300-P Secondary PRT with INFO-CON for 1523/1524 Models, 300 x 6mm, Calibration Options.