Fluke 5609-12-I Secondary PRT with Gold Pins, -200 to 670°C, Calibration Options

Model: 5609-12-I

This secondary PRT with gold pins has 4-wire connections and reference-grade platinum sensors.

This product has been discontinued. Click here for the replacement model(s).



Fluke 5609-12-I Offers

This secondary PRT with calibration options has a dry-well reference with a 90° bend.

Features

  • -328 to 1,238°F (-200 to 670°C) (3.94in (100mm) minimum immersion)
  • ±0.01 short-term stability
  • Calibration not included, NVLAP-accredited calibration optional, lab code 200348-0

Fluke 5609-12-I Specifications

Temperature Range -328 to 1,238°F (-200 to 670°C)
Nominal Resistance at 0.01°C 100Ω ±0.5Ω
Temperature Coefficient 0.0039250Ω/Ω/°C
Short-Term Repeatability ±0.01 at 0.010°C; ±0.02°C at max temp
Drift ±0.01 at 0.010°C; ±0.02°C at max temp
Hysteresis ±0.01°C maximum
Sensor Length 1.2in ±0.2in (30mm ±5mm)
Sensor Location 0.1in ±0.1in (3mm ±1mm from tip)
Minimum Insulation Resistance 500MΩ at 23°C, 10MΩ at 670°C
Transition Junction Dimensions 2.8 x 0.49in (71 x 12.5mm)
Minimum Immersion Length (<5 mK Error) 3.9in (100mm)
Maximum Immersion Length 12in (305mm)
Response Time 12 seconds typical
Self Heating (In 0°C Bath) 50mW/°C
Lead-Wire Cable Type Teflon, 24AWG
Lead-Wire Length 6ft (1.8m)
Lead-Wire Temperature Range -58 to 482°F (-50 to 250°C)
Calibration Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0; please see calibration uncertainty table and its explanation of changeable uncertainties
Click here for complete specifications on the Fluke 5609-12-I

What's included with the Fluke 5609-12-I

  • Fluke 5609-12-I Secondary PRT with Gold Pins, -200 to 670°C, Calibration Options