Fluke i6000s Flex-36 AC Current Probe

Part Number:
FE I6000S-FLEX-36
In Stock?
Please call for availability.

Quantity

Price:
$613.49 USD (change)

 

The Fluke i6000s Flex is an AC current probe utilizing the Rogowski principle. It can be used to measure currents up to 6000A when used in conjunction with oscilloscopes, recorders or data loggers. The flexible and lightweight measuring head allows quick and easy installation in hard to reach areas.
The Good:


The Bad:


The Bottom Line:

Full Review

Electrical Characteristics
Current Ranges

60 A / 600 A / 6000 A ACRMS
Output Sensitivity (AC coupled)

50 mV /5mV / 0.5mV per A
Load Impedance

100 kOhm minimum
Accuracy (at +25°C)

±1% of range (45 – 65 Hz)
Linearity (10% to 100% of range)

±0.2% of reading
Noise

8mV ACRMS (60 A)

2 mV ACRMS (600/6000A)
Frequency Range

10 Hz to 50 kHz (- 3 dB)
Phase Error

< ±1º (45 – 65Hz),

±10º (at 20 kHz)
Position Sensitivity (with cable > 25mm from the coupling)

±2% of range
External field (with cable > 200mm from the head)

±1% of range
Power Supply

2 x AA MN 1500 LR6 alkaline

400 Hours, low battery indicator or dedicated ext power supply
Temperature Coefficient

±0.08% of reading per C
Working Voltage (see Safety Standards section)

600 V ACRMS or DC

General Characteristics
Head Cable length…(double insulated)

610mm (24”) I6000s Flex-24
915mm (36”) I6000s Flex-36
Head Cable Diameter

14.3 mm
Head Bend Radius

38.1 mm
Cable length (head to electronics)

2 meters
Output Connection

0.5 m cable terminated with
safety BNC connector
Supplied with 4mm safety plug
adaptor
Operating Temperature Range

-20 to +90 °C (head)
-20 to +85 °C (electronics)
Storage Temperature Range

-40 to +105 °C (head)
-20 to +85 °C (electronics)
Operating Humidity

15% to 85% (non condensing)
Weight

180 g (head), 190 g (electronics)

Safety Standards
BS EN 61010

1: 2001
BS EN 61010

2-032: 2002
BS EN 61010

031: 2002
600VRMS, Category III, Pollution Degree 2


Note

Use of the probe on uninsulated conductors is limited to 600 V ACRMS or DC and frequencies below 1 kHz.
EMC Standards
BS EN 61326: 1998 +A1, A2, & A3